Hardness Test Blocks
PRODUCTS
HARDNESS TEST BLOCKS
ISO/DaKKS Certificated German Made Test Blocks
VICKERS HARDNESS REFERENCE BLOCKS
HARDNESS TEST BLOCKS
Hardness test blocks serve both for indirect testing and calibration of hardness testing machines, especially for daily verification of the hardness testing machines and thus they are indispensable for quality assurance in hardness testing to secure the correct tests. Our Vickers hardness reference blocks are available in the following sizes:
Vickers:30x30x6mm (Micro), 60x60x16mm (Macro)
Calibration: All hardness reference blocks are calibrated by a calibration laboratory according to DIN EN ISO 6507-3 or ASTM E92 and receive a DAkkS calibration certificate.


HIGH QUALITY OPTIC COMPONENTS
Equipped with very robust optomechanical system and top quality optic components (mirrors, half mirror, condenser lens).
COMPACT LED ILLUMINATION
It has compact LED illumination, completely hidden inside the frame.


TECHNICAL DATA
| ORDER NO | 72 01 |
| MODEL NO | M1 |
| Vickers HV as per ISO 6507, ASTM E384, E92 | HV0,01 - HV0,025 - HV0,05 - HV0,1- HV0,2 - HV0,3 - HV0,5 - HV1 |
| Knoop HK as per ISO 4545, ASTM E384, E92 | HK0,01 - HK0,025 - HK0,05 - HK0,1- HK0,2 - HK0,3 - HK0,5 - HK1 |
| Hardness conversions to other scales | As per ISO 18265 |
| Test Loads (gf) | 10gf – 25gf – 50gf – 100gf – 200gf - 300gf- 500gf - 1000gf |
| Test Loads (N) | 0,09807N - 0,2452N - 0,4904N - 0,9807N - 1,961N - 2,942N - 4,904N - 9,807N |
| Test Load Accuracy | 10gf - 200gf: +/- 1,5% and > 200gf: +/- 1,0% |
| As per ISO standard (and ASTM E-384, JIS B7725, JIS B7734) | |
| Test Load Dwell Time | from 5 - 99 sec with 1 sec increments |
| As per ISO standard (and ASTM E-384, JIS B7725, JIS B7734) | |
| Test Load Application Speed | 10gf - 200gf: < 0.070 mm/s and > 200gf: < 0.200 mm/s |
| As per ISO standard (and ASTM E-384, JIS B7725, JIS B7734) | |
| Load Force Application | Dead-weight, Automatic motorized, load, dwell, unload |
| Load Selection | Manual by external setting knob |
| Turret Configuration | Motorized 3-mount turret: 1 indenter, 2 objectives Position Auto-indication |
| Indenter | Including 1 Vickers Diamond Indenter (ISO/DaKKs certified, Made in Germany) |
| Objectives | Standard objectives 10X, 40X, optional 5X, 20X (Made in Germany) |
| Image capture by integrated CCD-camera (no eyepiece) | |
| Sample Image Capture | By Integrated CCD Camera, 5.0 MP, USB 3.0 (Made in Germany) |
| Indent Measurement Resolution | 0,00002 mm (0,02 µm) |
| Illumination Source | Built-in LED Light Source |
| Automatic Brightness Control | Standard for highest contrast and sharpness |
| Z-axis (Spindle Vertical Movement) | Manual |
| Auto-Focus AF | Not Available |
| Motorized XY-Stage | Not Available |
| (for Fully Automatic Multiple Test & Measurements) | |
| Travel Distance | Not Available |
| Controller | Built-in Windows operated mini-PC, communicates the hardness tester electronics by CANBUS Protocol |
| Interface Network | Interface network USB (x4), Displayport (x2), LAN (x1) and Wireless LAN |
| Auxilliary | Optional Keyboard, Mouse (wireless) |
| Monitor | External 22” Full HD Capacitive 10-point Multi-Touch Desktop Monitor, IPS Panel, Antistatic Shield, Tiltable, Integrated Power Adaptor |
| Operating System | Windows 10 |
| User Interface Software | METKON N-Sure |
| Test Settings | Scale, Load Dwell Time , Start, Turret positioning, Hardness Conversion, Illumination, Screen preference settings, Alerts for errors |
| Measurement Settings | Surface Image capture, Indentation capture, Manual measurement and Auto-measurement, Single indents, Manual pattern, Preset pattern, CHD, EHT, NHT, GO/NG judgement, Statistics, Conversion, Cilindrical correction, Language, Database, Test Reports with image analysis data, Data export, User management, Service and Calibration Settings |
| Operational Environmental Conditions | Ambient Operating Temperature Range: 23°C +/- 5°C |
| Ambient Humidity Range: RH 40% to 70% relative humidity, non-condensing | |
| Dimensions, WxDxH | 240x560x610 mm |
| Specimen Accommodation | Vertical 210 mm, Horizontal (throat depth) 165mm |
| Weight | 65 kgs |
| Mains | 100-240 V, 1-phase, 50/60 Hz, autosetting |
BROCHURE
GET YOUR BROCHURE FOR
HARDNESS TEST BLOCKS
Get the full introduction, descriptive visuals, and key feature descriptions to help you quali your choice of equipment.
• Explanatory photos
• Key feature descriptions
• Technical Data Sheet